GW INSTEK GDS-3000 Series

Features

* 500/350/250/150MHz with 2/4 Channels
* 5GSa/s RT or 100GSa/s ET Sampling Rate
* Independent Memory for Each Channel 
* VPO Technology 
* Large 8-inch 800x600 Display 
* Split Screen Function 
* 3 Built-in Impedances (50Ω/75Ω/1MΩ)
* Power Analysis Software (Optional)
* Serial Bus Analysis Software for I2C, SPI and UART (Optional)

 

 A Hi-tech DSO Platform
    The GDS-3000 Series is a new platform of 4-input channels, 500MHz bandwidth, 5GSa/s sampling rate, and VPO waveform display. The split screen feature has been designed to meet the requirements of multi-window & multi-signal tests in the research and the manufacturing fields. The optional power analysis software and the optional serial bus analysis software are available to facilitate the engineer's tasks in testing and manufacturing of the associated products.
    Three new differential probes, GDP-025, GDP-050 & GDP-100, and two new current probes, GCP-005 ,GCP-020 ,GCP-100 and GCP-530 & GCP-1030, are coming along with the GDS-3000 Series to provide total solutions for a wide variety of applications in the industry, service and education market sectors. The GDS-3000 Series, a high-tech platform carrying thoughtful features, brings very high customer value to both general purpose market and professional market.

 

 

VPO- Visual Persistence Oscilloscope Signal Processing Technology 


UNIQUE SPLIT SCREEN FUNCTION

    The unique split screen feature of GDS-3000 Series allows each input channel to be operated independently with respective setting and waveform display. The time base, the vertical sensitivity, and the trigger selections can be done by each channel separately, and the waveform of each input signal can be shown on the individual part of the screen. This nearly four-DSO-in one feature is very useful for the applications that need to simultaneously see the details of multiple waveforms with very different characteristics. The 8-inch high resolution 800x600 LCD display makes the split screen a pleasant observation environment to view the details of complex signals.


COMPLETE SET of TRIGGER FUNCTIONS

    Besides Edge trigger, the GDS-3000 Series also offers various trigger functions, including Video, Pulse Width, Runt, Rise Time & Fall Time (specific time length), Alternate, Delay by Time, Delay by Event, and Hold-Off. The high sampling rate, the VPO signal processing & display, and the flexible trigger function all together make the GDS-3000 Series a powerful tool for waveform capture and display of various types of signals.


AUTO RANGE FOR BOTH TIME BASE AND VERTICAL SCALE
Autorange
    The Auto Range function automatically adjusts the time base and/or the vertical scale of displayed waveform when the frequency and/or the amplitude of input signal changed. This function gives user the convenience to have DSO always display waveform in a proper fashion on the screen tracking the frequency and amplitude changes of the input signal. It is especially useful when the user needs to alternately probe and test multiple circuit points containing signals with different frequencies and amplitudes.
 

DUAL DISPLAY WINDOW ZOOM
WindowZoom     The GDS-3000 Series Window Zoom function provides dual display mode to show the main waveform and the magnified section of zoomed-in waveform at the same time. Under "Zoom" mode, the width and the position of zoom-in window over the main waveform can be selected to get the magnified waveform as needed for detailed observation. To quickly and accurately move the zoom-in window to the expected position, the "Coarse" mode helps move the window to the needed position immediately and the "Fine" mode provides fine adjustment to precisely place the window in the exact position.
 
28 AUTOMATIC MEASUREMENTS
Automeasurement
    The GDS-3000 Series supports simultaneous measurement of up to 28 waveform measurement items grouped into three main waveform parameters: amplitude, time and delay measurements. The display modes include an individual mode and a Display All mode. The former can display any 8 of the automatic measurements while the later can display all the automatic measurements for a channel.
 

FFT TEST FUNCTION
FFT     To observe fundamental and harmonic frequency components of a signal, the FFT function on a digital storage oscilloscope is often used. Typically the traditional unit of the FFT is decibel (dB). However, when using dB it is sometimes difficult to identify the fundamental frequency of a signal from a noisy spectrum. With FFTrms function, the GDS-3000 Series can clearly display the fundamental frequency of an acquired waveform. 
    The FFT function of GDS-3000 supports Rectangular, Hamming, Hanning, and Black-harris windows.
 
THREE INPUT IMPEDANCE SELECTIONS
Impedance
    Three input impedance, 1M , 75 , and 50 are available for user's selection. The flexibility of impedance selections, including 1M to get minimum loading effect, 75 to accommodate Video transmission applications and 50 to fit RF communication applications, extends the GDS-3000 Series utilization range.
 

X-Y MODE
 
    The X-Y mode of GDS-3000 defines CH1 and CH3 as the horizontal axis and CH2 and CH4 as the vertical axis, allowing the display of 2 sets of X-Y pattern simultaneously. The measurement items include Rectangular, Polar, Product and Ratio that fits most of the popular X-Y applications.
    The X-Y pattern and the time domain waveforms can be shown on the screen simultaneously. Two cursors on the time domain waveforms allow the identification of cursor-associated locations on the X-Y pattern display. 
 
 
EXTENDABLE APPLICATION SOFTWARE
    The GDS-3000 Series allows future installation of additional application software at the user site. This provides an open environment for optional software upgrade and additional feature built-in in whenever the GDS-3000 Series user has the need. The flexibility of software installation platform keeps the DSO being in use always up-to-date.
 
WAVEFORM FILE PREVIEW
    The GDS-3000 provides an optimized operation interface for viewing screen captures. Generally, the oscilloscope may store large amounts of waveform data after a long period of time. To help prevent engineers from selecting the wrong file from a large number of stored waveform files, the screen capture preview function can be used to preview the waveform file without opening files so that operation of the oscilloscope is more efficient and convenient.
 
 
FREE REMOTE CONTROL SOFTWARE

    Using a USB port coupled with FreeWave remote monitoring software is the easiest and most convenient way to capture data from the GDS-3000 Series. With FreeWave, a screenshot can be saved as an image file (.bmp/.jpg), waveform data (.csv) can be logged and movie files(.wmv) can be recorded in real-time. Not only can FreeWave monitor and record waveforms over a long period of time, but previously recorded waveforms can also be observed. Instrument settings can even be configured without the need to learn incomprehensible command line syntax. With the simple user interface and robust features, FreeWave allows you to get the most out of the GDS-3000 with little effort.
 

SVGA OUTPUT

SVGA

    A SVGA video output port in the rear panel of GDS-3000 Series allows the screen-image transfer from DSO to an external projector or a monitor for remote monitoring or big screen observation. This direct image transfer feature greatly increase the efficiency of presentation in the meeting, teaching in the class, remote monitoring of hazardous events from a secured zone, and fast and easy monitoring in the production line.
 

VARIOUS INTERFACES SUPPORT
Interface
    Two high-speed USB 2.0 Host ports located in both front panel and rear panel are used for easy access of stored data. In the rear panel, a USB Device port is available for remote control and hardcopy print-out through a PictBridge compatible printer. RS-232 and LAN interfaces are provided as standard for system communication & ATE applications.
A SVGA video output port allows the transfer of DSO screen image to an external projector or monitor for remote monitoring or big screen observation. A GPIB to USB adaptor is available as an option for interface conversion though the USB Device port in the front panel.
 

SERIAL BUS ANALYSIS SOFTWARE SUPPORTING I2C, SPI and UART (OPTIONAL)
Serial Bus Analysis Software
The GDS-3000 Series provides two dedicated keys
 
    With serial bus technology being widely used in embedded applications, the proper triggering and analysis of flowing data, control signal and associated pulse waveforms in serial bus  communication has been a difficult job and challenge to design engineers. The Serial Bus Analysis software of GDS-3000 Series carries complete analysis tools for triggering and decoding of commonly used serial bus interfaces, including I2C, SPI and UART. Without spending time to study serial bus regulation details, the user only needs to set the trigger condition on GDS-3000 to get the data slots of interest.
 

POWER ANALYSIS SOFTWARE FOR POWER SUPPLY MEASUREMENTS (OPTIONAL)
Power Quality
Harmonics
   
Ripple
In-rush Current
 
    The Power Analysis software contains four measurement functions, including Power Quality, Harmonics, Ripple and Inrush Current. The Power Quality analysis function allows the measurements of Voltage, Current, Frequency, Power and other quality related parameters for power source efficiency improvement. The Harmonics analysis function performs evaluation of power waveform distortion and gives harmonic test data for power source design and quality check. This function is complied with IEC 61000-3-2 standard. The Ripple measurement function, acquiring the ripple and noise overriding the DC waveform, is used to evaluate the DC power source quality. The Inrush Current measurement function is used to measure the power-on surge current, which may cause the damage of the device circuit.

Specification:

  GDS-3152 GDS-3154 GDS-3252 GDS-3254 GDS-3352 GDS-3354 GDS-3502 GDS-3504
Vertical Sensitivity  
Input Channels 2Ch+EXT 4Ch+EXT 2Ch+EXT 4Ch+EXT 2Ch+EXT 4Ch+EXT 2Ch+EXT 4Ch+EXT
Bandwidth DC~150MHz(-3dB) DC~250MHz(-3dB) DC~350MHz(-3dB) DC~500MHz(-3dB)
Rise Time 2.3ns 1.4ns 1ns 700ps
Vertical Resolution 8 bits
Vertical Resolution @ 1MΩ 2mV~5V/div
Vertical Resolution @ 50/75Ω 2mV~1V/div
Input Coupling AC, DC, GND
Input Impedance 1MΩ// 15pF  
DC Gain Accuracy ±(3% X |Readout| + 0.1div + 1mV)
Polarity Normal, Invert
Maximum Input Voltage @1MΩ 300Vrms, CAT I(300Vrms CAT II with GTP-151R/251R/351R/501R 10:1 probe  )
Maximum Input Voltage @50/75Ω 5 VRMS max, CAT I
Offset Position Range  2mV/div ~ 100mV/div : ±0.5V ;200mV/div ~ 5V/div : ±25V
Bandwidth Limit GDS-3152/3154:20MHz
(-3dB)
GDS-3252/3254: 20MHz/100MHz (-3dB) GDS-3352/3354: 20MHz/100MHz/
200MHz  (-3dB)
GDS-3502/3504: 20MHz/ 100MHz/ 200MHz/ 350MHz (-3dB)
Waveform Signal Process Add, subtract, multiply, and divide waveforms, FFT, FFTrms
  FFT: Spectral magnitude.
Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning or Blackman-Harris.Integration*, Differentiation* *: App installation required
Trigger  
Source 2 CH model: CH1, CH2, Line, EXT
4 CH model: CH1, CH2, CH3, CH4, Line, EXT
Trigger Mode Auto (supports Roll Mode for 100 ms/div and slower), Normal, Single Sequence
Trigger Type Edge, Pulse Width(Glitch), Video, Runt, Rise & Fall(Slope), Alternate, Event-Delay(1~65,535 events), 
Time-Delay(Duration ,10ns~10s),
I²C, SPI, UART(optional)
  Runt: Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again.
  I²C (optional)
Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data, or Address and Data on I²C buses 
  SPI (optional)
Trigger on SS, MOSI, MISO, or MOSI and MISO on SPI buses 
  UART (optional)
Trigger on Tx Start Bit, Rx Start Bit, Tx End of Packet, Rx End of Packet, Tx Data, Rx Data, Tx Parity Error, and Rx Parity Error
Trigger Holdoff Range  10ns~10s
Coupling AC, DC, LF rej., Hf rej., Noise rej.
Sensitivity DC ~ 50MHz Approx. 1div or 10mV
50MHz ~ 350MHz Approx. 1.5div or 15mV
EXT Trigger  
Range ±15V
Sensitivity DC ~ 150MHz  Approx. 100mV ;150MHz ~ 250MHz  Approx. 150mV 
250MHz ~ 350MHz  Approx. 150mV ;350MHz ~ 500MHz  Approx. 200mV 
Input Impedance 1MΩ±3%, ~16pF
Horizontal  
Time base range 1ns/div ~ 100s/div (1-2-5 increments;GDS-3502/3504:1-2.5-5 increments); ROLL : 100ms/div ~ 100s/div
Pre-trigger 10 div maximum
Post-trigger 1,000 div
Time base accuracy ±20 ppm over any ≧1 ms time interval
X-Y Mode  
X-Axis Input Channel 1; Channel 3
Y-Axis Input Channel 2; Channel 4
Phase Shift ±3° at 100kHz 
Signal Acquisition  
Real Time Sample Rate 2.5GSa/s 5GSa/s 2.5GSa/s 5GSa/s 5GSa/s 5GSa/s 4GSa/s 4GSa/s
ET Sample Rate 100GSa/s maximum for all models
Record Length 25k points/ channel
Acquisition Mode Normal, Average, Peak detect, High resolution, Single
  Normal: Acquire sampled values.
  Average: From 2 ~256 waveforms included in average.
  Peak Detect: Captures glitches as narrow as 2 ns at all sweep speeds
  Hi Res: Real-time boxcar averaging reduces random noise and increases vertical resolution.
Cursors and Measurement  
Cursors Amplitude, Time, Gating available
Automatic Measurement  28 sets: Vpp , Vamp , Vavg , Vrms , Vhi , Vlo , Vmax , Vmin , Rise Preshoot/ Overshoot , Fall Preshoot/Overshoot, Freq , Period , Rise Time , Fall Time , Positive Width , Negative Width , Duty Cycle, Phase angle, and eight different delay measurements (FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF)
Cursors measurement Voltage difference between cursors ( ∆V) Time difference between cursors ( ∆T)
Auto counter 6 digits, range from 2Hz minimum to the rated bandwidth
Power Measurements (option)  
Power Quality VRMS, VCrest Factor, Frequency, IRMS, ICrest Factor, True
Measurements Power, Apparent power, Reactive power, Power factor, Phase angle.
Harmonics Freq, Mag, Mag rms, Phase, THD-F, THD-R, RMS
Ripple Measurements V ripple ,I ripple 
In-rush current First peak, Second peak
Control Panel Function   
Autoset Single-button, automatic setup of all channels for vertical, horizontal and trigger systems, with undo autoset
Auto-Range Allow users to quickly move from test point to test point without having to reset the oscilloscope for each test point
Save Setup 20 set
Save Waveform 24 set
Display System  
TFT LCD Type 8" TFT LCD SVGA color display
Display mode YT ;XY
Display Resolution 800 horizontal × 600 vertical pixels (SVGA)
Interpolation Sin(x)/x & Equivalent time sampling
Waveform Display Dots, vectors, variable persistence, infinite persistence
Display Graticule 8 x 10 divisions
Display Brightness Adjustable
Waveform capture rate
3500 waveform /sec real time
Interface  
RS-232C DB-9 male connector
USB Port  2 sets USB 2.0 High-speed host port ; 1 set USB High-speed 2.0 device port 
Ethernet (LAN)Port  RJ-45 connector, 10/100Mbps
SVGA Video Port  DB-15 female connector, monitor output for display on SVGA monitors
GPIB GPIB to USB Adapter (GUG-001 Option)
Go/NoGo BNC 5V Max/10mA TTL Open collector output
Internal flash disk 64MB
Kensington Style Lock Rear-panel security slot connects to standard Kensington-style lock.
Line output 3.5mm stereo jack for Go/NoGo audio alarm
Power Source  
Line Voltage range AC 100V ~ 240V, 48Hz ~ 63Hz, Auto selection
operation environment  Temperature: 0°C to 50°C. Relative Humidity ≤ 80% at 40°C or below; ≤ 45% at 41°C ~ 50°C
Miscellaneous  
Multi-language menu Available
On-line help Available
Time clock Time and Data, Provide the Data/Time for saved data
Dimensions & Weight 400(W) X 200(H) X 130(D)mm, approx 4 kg

User manual *1, Power code *1,


GTP-151R: 150MHz(10:1) passive probe for GDS-3152/ 3154 (per channel)


GTP-251R: 250MHz(10:1) passive probe for GDS-3252/ 3254 (per channel)

GTP-351R: 350MHz(10:1) passive probe for GDS-3352/ 3354 (per channel)

GTP-501R: 500MHz(10:1) passive probe for GDS-3502/ 3504 (per channel)

Optional Accessories:

GTC-001 Instrument cart,450(W) x 430(D)mm (120V input socket)
GTC-002 Instrument cart 330(W) x 430(D)mm (120V input scoket)
GTL-110 Test lead, BNC to BNC connector
GTL-232 RS-232C cable, 9-pin female to 9-pin female, Null Modem for computer
GTL-242 USB 1.1 cable, A-B type cable 4P, 1800mm
GTP-033A 35MHz 1:1 passive probe
GTP-352R 350MHz(20:1) passive probe
Differential Probe GDP-025: 25MHz high voltage differential probe
GDP-050: 50MHz 25MHz high voltage differential probe
GDP-100: 100MHz 25MHz high voltage differential probe
Current Probe

GCP-020 :10KHz/200A current probe

GCP-100 :100KHz/100A current probe

GCP-530: 50MHz/ 30A current probe
GCP-1030: 100MHz/ 30A current probe
GCP-206P: power supply for current probe (2 input channels)
GCP-425P: power supply for current probe (4 input channels)

 

 

Ordering Information:

GDS-3152 150MHz, 2 channel, color LCD display DSO
GDS-3154 150MHz, 4 channel, color LCD display DSO
GDS-3252 250MHz, 2 channel, color LCD display DSO
GDS-3254 250MHz, 4 channel, color LCD display DSO
GDS-3352 350MHz, 2 channel, color LCD display DSO
GDS-3354 350MHz, 4 channel, color LCD display DSO
GDS-3502 500MHz , 2 channel,  color LCD display DSO
GDS-3504 500MHz , 4  Channel ,  color  LCD  display  DSO


Option:

GUG-001 GPIB to USB adapter
DS3-SBD Serial bus analysis software I2C / SPI/ UART (for 4 channel modelonly)
DS3-PWR Power analysis software: Power quality/Harmonic/Ripple/In-rush current measurement

 

 